Crest Factor Measurement with the 2700 Series

Created 24 Feb 2009

Question:

How can I measure the crest factor of a waveform on a 2722 analyzer?

Answer:

Crest factor is the difference between the Peak and RMS values of a signal. To measure it, we will need to put the same signal into both Channel A and Channel B analyzer inputs. If we are measuring an internally generated signal, we can set the inputs to GenMon. If we are measuring an external signal, we’ll need two identical signals or else a Y adapter.

Now, set the Function Meter to 2-Ch. Ratio with dB units, and set the Detector (Det) to Peak. The meter can be set to measure either channel A or B—it doesn’t matter. Now, the Function Meter will be reading the peak value on the selected channel and comparing it to the RMS value measured by the Level Meter on the other. The resulting display is the Crest Factor.

SLX #4154
Eric


Related Downloads

2700 Series Crest Factor test

File configured to measure crest factor. See knowledge base article for complete explanation.

CrestFactor.at27 (67 KB)

Return to Knowledge Base home

Comments

2 comments so far
Dan Dickey said this 10 months ago ...
weighted noise measurement using RMS detector
Dan Dickey said this 10 months ago ...
QPK detector

Please log in to add a comment.

Back To Top

Accredited by A2LA

Browse Products View Solutions Schedule a Demo Get Service Register

Audio Precision

Copyright © 2009 Audio Precision Inc. All Rights Reserved.    |  Privacy Policy